The professional body for laser safety in the United Kingdom
Guidance for clinics

Laser and IPL Local Rules

Local Rules are the written safety procedures for using a Class 3B or Class 4 laser or IPL system in a specific clinic. They are drafted by the Laser Protection Adviser, adopted by the employer and signed by every authorised user. They set out who is responsible, the controlled area, protective eyewear, key control, pre-use checks and what to do if something goes wrong.

Last checked 21 October 2026

What Local Rules are

Local Rules turn the general duties in health and safety law into instructions for one set of premises. The Control of Artificial Optical Radiation at Work Regulations 2010 require employers to assess exposure to laser and IPL light and to keep it below the exposure limits. The Management of Health and Safety at Work Regulations 1999 require a suitable and sufficient risk assessment. Local Rules record the controls that follow from that assessment, in the order staff need them.

The MHRA 2015 guidance on lasers, intense light source systems and LEDs describes Local Rules as part of the safety structure every clinic using Class 3B or Class 4 lasers or IPL should have, alongside a Laser Protection Adviser, a Laser Protection Supervisor and a register of authorised users. A summary of each source is on the standards and legislation page.

Who writes and signs them

The Laser Protection Adviser drafts the Local Rules after assessing the treatment rooms and devices. The employer, as duty holder, adopts them and remains legally responsible for the safe use of the equipment. The employer may delegate tasks to the adviser and the supervisor, but not that responsibility.

  • Laser Protection Adviser. Writes the rules, advises on eyewear and the controlled area, and reviews the rules at each audit.
  • Laser Protection Supervisor. A member of the clinic’s staff who makes sure the rules are followed day to day and keeps the records.
  • Authorised users. Every person allowed to operate a device reads the rules, has them talked through, and signs a declaration.

Certified advisers are listed on the Register of Laser Protection Advisers.

What Local Rules must contain

The MHRA 2015 guidance sets out what Local Rules should cover. Use this list to check a clinic’s current document.

  • The safety structure: the employer, the Laser Protection Adviser and the Laser Protection Supervisor, with contact details
  • A description of each laser or IPL device covered, including its class and wavelength
  • The register of authorised users for each device
  • The boundaries of the laser controlled area and how access to it is controlled
  • Warning signs and the entry arrangements while the device is in use
  • Key control: who holds the key and where it is kept when the device is not in use
  • Protective eyewear for operators, clients and anyone else in the room, matched to the wavelength in use
  • Pre-use safety checks and the normal operating procedure
  • What to do in an adverse incident or suspected eye exposure, and how it is reported
  • Arrangements for servicing and for visiting engineers
  • A signed declaration from every authorised user that they have read and understood the rules

Each item should name the actual rooms, devices and people. Rules that refer to “the laser” without a model, wavelength or room are a common audit finding.

MHRA 2015 guidance (PDF) (opens in a new tab)

Sample contents page

The section headings of the Association’s Local Rules template, which certified advisers complete for each clinic. Headings only: the content under each one is written for the premises.

  1. Purpose of the Local Rules
  2. 1. Organisation
  3. 2. Devices used at this clinic
  4. 3. Laser controlled area
  5. 4. Potential hazards to persons
  6. 5. Expert treatment protocol
  7. 6. Client safety
  8. 7. Register of authorised users
  9. 8. Laser Protection Supervisor
  10. 9. Laser user and staff training
  11. 10. Record of laser use
  12. 11. Responsibilities of authorised users
  13. 12. Equipment identification and safety labelling
  14. 13. Warning signs
  15. 14. Security and custody of the laser key
  16. 15. Laser protective eyewear
  17. 16. Personal protective equipment
  18. 17. Servicing and maintenance
  19. 18. Reflective surfaces
  20. 19. Adverse incidents
  21. 20. Pre-use safety checks
  22. 21. Normal operating procedure
  23. 22. Safe system of work
  24. 23. Fire safety
  25. 24. Equipment faults
  26. 25. Emergency shutdown procedure
  27. 26. Prevention of unauthorised use
  28. 27. Duties of the Laser Protection Supervisor
  29. 28. Maintenance procedures

When to review and update them

Local Rules should be reviewed and re-signed whenever any of the following changes:

  • a new laser or IPL device, or a change of handpiece or wavelength
  • a new treatment room, or changes to an existing room such as windows, mirrors or doors
  • a new treatment or procedure
  • a change of Laser Protection Adviser or Laser Protection Supervisor
  • new authorised users, who sign before they operate the equipment
  • an adverse incident or near miss

The Association’s template also sets a formal review at least once a year. Under the MHRA guidance the supervisor reviews the risk assessment every year and the adviser audits it at least every two years. The laser safety audit covers both documents.

How councils and regulators check them

England. Cosmetic laser and IPL premises are mostly regulated through local authority licensing, and conditions differ by council. In London, premises usually need a special treatment licence. Licence conditions commonly require Local Rules written by the Laser Protection Adviser and available in the treatment room. Where treatment is a regulated medical activity, the provider registers with the Care Quality Commission.

Scotland. Independent clinics are regulated by Healthcare Improvement Scotland.

Wales. Providers of Class 3B and Class 4 laser and IPL treatments register with Healthcare Inspectorate Wales.

Northern Ireland. Establishments using Class 3B and Class 4 lasers or IPL register with the Regulation and Quality Improvement Authority. Inspection reports routinely review the adviser’s Local Rules and risk assessment.

Inspectors typically look for a current, signed copy in the treatment room, rules that match the devices on site, and signatures from every authorised user. More on each nation is on the laser safety for clinics page.

Local Rules, risk assessment and treatment protocols

The three documents are often confused. Each has a different job.

DocumentPurposeUsually written by
Risk assessmentIdentifies the hazards of each device and room, who is at risk and the controls needed.Laser Protection Adviser with the employer
Local RulesSets out the safety procedures staff follow to apply those controls.Laser Protection Adviser
Treatment protocolsClinical instructions for each treatment: suitability, settings, patch tests and aftercare.A suitably qualified clinician

The risk assessment comes first; the Local Rules follow from it. See laser and IPL risk assessment for what that document covers.

Questions

Are Local Rules a legal requirement?

The MHRA 2015 guidance has no legal status in itself, but the Control of Artificial Optical Radiation at Work Regulations 2010 require employers to control exposure, and written Local Rules are the standard way of showing those controls. Council licence conditions and inspectors commonly ask to see them.

Who should write Local Rules for a laser or IPL clinic?

The Laser Protection Adviser normally drafts them for the specific premises and devices. The employer or registered manager adopts them, and the Laser Protection Supervisor makes sure they are followed each day.

Can I use a laser local rules template?

A template is a starting point only. Local Rules must describe the actual rooms, devices, wavelengths, eyewear and staff of the clinic, so a generic document downloaded from the internet will not meet the guidance without being rewritten for the premises.

Does each device need its own Local Rules?

Each device must be covered. One document can cover several devices in the same clinic, provided it describes each one, its controlled area, its eyewear and its authorised users.

How often should Local Rules be reviewed?

Whenever something changes: a new device, a new room, a new procedure, new staff or a new adviser. The Association’s template also sets a formal review at least once a year, and staff re-sign the rules after each update.

Does my laser protection advisor need to visit to write them?

The guidance expects Local Rules to reflect the actual premises, including the controlled area, windows, reflective surfaces and access. In practice that means the adviser assesses the rooms before the rules are finalised.

Who has to sign the Local Rules?

Every authorised user signs a declaration that they have read and understood them. The rules are also talked through with staff, and new starters sign before they operate the equipment.

Book your LPA audit

A certified Laser Protection Adviser audits your clinic. You receive Local Rules, a risk assessment, an audit report and a Certificate of LPA Audit valid for 24 months.